3D.6.0 AEC CALIBRATION (WALL BUCKY)
WALL
BUCKY
Please note the following points regarding wall Bucky calibration:
If the wall Bucky is dedicated to chest radiography, a focused grid with a 10:1 or 12:1 ratio should be used along with an SID of 72 in. (180
cm).
If the wall Bucky will be used for conventional as well as chest radiography, then two grids should ideally be used. See the note on the next
page.
A reasonable compromise if a single grid must be used is a 10:1 ratio, 60 in. (150 cm) grid.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
Service Manual
CTM
Absorber
O.T.
FRONT VIEW OF
WALL BUCKY
72 in(180cm)
TYPICAL WALL BUCKY IMAGE RECEPTOR
Figure 3D-8: Equipment setup for wall Bucky AEC calibration
Ch # 901476-05
Rev. A
3D
AEC Calibration
Page 3D-31