3C.11.3 DAP Calibration Procedure (Cont)
Step
15.
Calculate the percentage error between the manually calculated DAP measurement (table 3C-4, step 3) and the measured
DAP value (table 3C-4, step 4). Record the percentage error in step 5 of the table.
16.
If the displayed DAP reading does not meet the required accuracy, increase or decrease CAL. VALUE 1 (membrane
console) or Calibration Value (DAP Device Type 1) for GenWare® by the same percentage as the percentage error.
17.
Repeat steps 7 to 18 until the required accuracy is obtained. Make as many copies of table 3C-4 as required to record the
results from all required iterations.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-78
X-RAY CASSETTE
ILLUMINATED
COLLIMATOR
FIELD
Figure 3C-21: Irradiated area vs. available image area
CMP 200
Action
13.2
EXPOSED
13.6
AREA
DEVELOPED FILM
®
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C