DP6000 Process Meter

SPECIFICATIONS

Except where noted all specifications apply to operation at +25°C.

General

DISPLAY
DISPLAY
INTENSITY
DISPLAY
UPDATE RATE
OVERRANGE
UNDERRANGE
DISPLAY
ASSIGNMENT
PROGRAMMING
METHODS
NOISE FILTER
FILTER BYPASS
RECALIBRATION All ranges are calibrated at the factory. Recalibration is
MAX/MIN
DISPLAY
PASSWORD
POWER
OPTIONS
ISOLATED
TRANSMITTER
POWER SUPPLY
NON-VOLATILE
MEMORY
FUSE
Main display: 0.60" (15 mm) high, red LEDs
Second display: 0.46" (12 mm) high, red LEDs
6 digits each (-99999 to 999999), with lead zero blanking
Eight user selectable intensity levels
5/second (200 ms)
999999
Display flashes
-99999
Display flashes
The main (Big) and small (Little) displays may be assigned
to PV1, PV2, PCT, d r-u, d gross, d nt-g, max/min, max &
min, set points, units (small display only), or Modbus input.
Four front panel buttons, digital inputs, PC and DP6000-SOFT
software, Modbus registers, or cloning using Copy function.
Programmable from 2 to 199 (0 will disable filter)
Programmable from 0.1 to 99.9% of calibrated span
recommended at least every 12 months.
Max/min readings reached by the process are stored until
reset by the user or until power to the meter is turned off.
Three programmable passwords restrict modification of
programmed settings.
Pass 1: Allows use of function keys and digital inputs
Pass 2: Allows use of function keys, digital inputs and editing set/reset points
Pass 3: Restricts all programming, function keys, and digital inputs.
85-265 VAC 50/60 Hz, 90-265 VDC, 20 W max
or jumper selectable 12/24 VDC  10%, 15 W max
Terminals P+ & P-: 24 VDC  10%. 12/24 VDC powered models
selectable for 24, 10, or 5 VDC supply (internal jumper J4).
85-265 VAC models rated @ 200 mA max, 12/24 VDC powered
models rated @ 100 mA max, @ 50 mA max for 5 or 10 VDC
supply.
All programmed settings are stored in non-volatile memory
for a minimum of ten years if power is lost.
Required external fuse: UL Recognized, 5 A max, slow
blow; up to 6 meters may share one 5 A fuse
10