7
Camera and Lighting Calibration
Lighting Calibration (Manual)
7-10
From the Tools menu, click Lighting Controller under the
1
Camera folder to display the following window.
A lighting ring is enabled when the check box beside the ring
name is selected. When a lighting ring is enabled, the
corresponding current in the mA/4 column is sent to the
lighting controller. Each lighting ring has a maximum
current; in addition, the current that may be sent to the
lighting controller must not exceed 4000 mA.
2
Ensure that the current for each enabled lighting ring is set
to the correct value and complies with the value set in the
Targets column.
To undo changes made to the output values, click Undo.
These values also have to be logged by clicking Log.
3
The camera will then take one image with each bank of LEDs,
and puts the results in the file Intensity.log, located in the
C:\Cpi\Log folder.