Testing the Move, Stop, and Jog Functions
Agilent U1840A Medalist sj5000 Hardware Manual
To test the Move function, follow the instructions below:
Enter 200 mm in the Required Width field and click Move.
1
2
Wait until 200 mm is displayed in Current Width.
3
Using Vernier calipers, check that the width of the conveyor
is 200 mm ±0.5 mm at both ends.
Enter 100 mm into the Required Width field and click Move.
4
5
Wait until 100 mm is displayed in Current Width.
6
Using Vernier calipers, check that the width of the conveyor
is 100 mm ±0.5 mm at both ends.
7
Exit the AWA menu, wait 5 seconds, and re-enter the AWA
menu.
8
Enter 150 mm into the Required Width field. The conveyor
should first home and then move to 150 mm.
9
Using Vernier calipers, check that the width of the conveyor
is 150 mm ±0.5 mm at both ends.
To test the Stop function, do the following:
Enter 300 mm in the Required Width field and click Move.
1
Before the conveyor reaches this width, click Stop.
2
3
Using calipers, measure the distance between the two rails
and check that it corresponds to the value in Current Width.
To test the Jog function, do the following:
Select a step size and click Jog Out or Jog In to jog the
1
conveyor rails.
The value in the Current Width field should increase or
decrease correspondingly.
2
Using Vernier calipers, measure the distance between the
two rails and check that it corresponds to the value in
Current Width.
3
This test should be run for all three conveyor step sizes, for
both the Jog Out and Jog In buttons.