Maintenance
Quarterly Maintenance
WA RNING
12-8
Carry out the following maintenance procedure every quarter:
• Carry out the monthly routine.
• When inspecting the control panel, check the tightness of
several connections at random. Ensure that no wires in the
terminals have become loose.
• Push the XY stage to one end of its travel.
• Remove any accumulated dust or debris from inside the
assembly.
• Vacuum clean the base of the table.
• Remove any dirty or dried lubricant from the bearing guides.
• Using a clean cloth, wipe along the rails to clean the bearing
guides. A cotton swab soaked in a general purpose white
spirit will suffice to remove stubborn debris.
• After the solvent has evaporated (if used), apply a thin,
continuous film of lubricant to the bearing guides. A good
quality natural bristle artist's brush is an excellent
applicator for this.
• Using a grease gun with a grease nipple (type UU), lubricate
the four bearing trucks on each gantry axis.
Failure to lubricate the stage at quarterly intervals may cause
damage to the rails and invalidate all warranties.