Power and Communications

Electrical Power
WA RNING
WA RNING
Agilent U1840A Medalist sj5000 Hardware Manual
Table
Table A-3
in the appendix lists the power specifications
for the sj5000.
The equipment must be installed according to the applicable
national wiring code (NEC) and all applicable local regulations.
All system sub-circuits are 100–120 VAC and 200–240 VAC. Only
connect equipment rated to this voltage to the system.
• The main incoming protected earth must be terminated at
the earth terminal.
• The earth link wire must have a minimum cross-sectional
area of 4 mm² (or 10 AWG).
• An uninterruptible power supply (UPS) is optional. Ensure
that the input voltage indicated on the UPS is same as the
supply voltage.
MAINS VOLTAGE DEPENDS ON THE UPS USED. BEFORE
CONNECTING TO THE MAINS, ENSURE MAINS VOLTAGE IS THE
SAME AS THE UPS INPUT VOLTAGE
100–120 V or 200–240 V
DETAILS ON MACHINE PLATE
Introduction to the Medalist sj5000
2
2-13