Camera and Lighting Calibration
Auto Calibration
7-12
The auto calibration feature requires the use of a calibration
plate (optional, not included with the sj5000). Before starting
the auto calibration process, you must set the location of the
calibration plate.
1
Ensure that there are no boards in the sj5000.
From the Tools menu, click Lighting Controller under the Camera
2
folder.
Click Set Location.
3
The system turns the LEDs on and displays the following
instruction:
Jog the camera as instructed and click Set.
4
Click Yes to confirm.
5
The system is now ready for lighting auto calibration.
Click Auto Calibration.
6
The camera automatically moves to the location of the
calibration plate and calibration will start.
7
You will be informed when the calibration is successfully
completed. Click OK.
The current outputs for each ring before and after calibration
are shown. These values area also written to the file
LightingController.log.