N O T E
Agilent U1840A Medalist sj5000 Hardware Manual
6
Check the Kollmorgen drives. The display on the drive should
show 2. This means that the drive has enabled correctly.
Make sure the axes are in the middle of the machine.
7
Go to the MS Summary window. On the Actions tab click the
Move to Further button. The axis should move in a positive
direction. Click Move to Further again and the axis should
move back.
11
Troubleshooting
11-35