Agilent Technologies U1840A Medalist sj5000 Hardware Manual page 165

Automated optical inspection system
Table of Contents

Advertisement

Agilent U1840A Medalist sj5000 Hardware Manual
As the Y axis is moved to the left, the numbers highlighted
above should increase.
c
If feedback in the opposite direction is received, then the
sj5000.xmp file has to be downloaded and saved to the
MEI card again. (The encoder phase should be reversed
for all motors, see the Motor Summary window.)
d
Make sure the both axes are in the middle of the machine.
Close the front access door and reset the E-Stop circuit.
e
Enter the following values in the fields:
On the MoCon toolbar, click the F button.
3
On the Filter tab, make sure both Filter 0, Controller 0 and
a
Filter 1, Controller 0 appear in the Object List. Click OK.
11
Troubleshooting
Encoder resolution
is 1 um, so this
means a travel
distance of 60 mm
(6000 x 1 um)
11-31