8
Camera and Lighting Controller Testing
8-4
Follow these steps to check the camera settings.
1
In Windows Explorer, navigate to C:\cpi\plc\terminal and
double-click terminal.exe.
2
From the Terminal window, open dalsa.trm.
3
Type
in the command line.
GCP
4
Check that SYNC Frequency, Internal Exposure Time,
Camera Link Mode, and Video Mode are as shown in the
following: