Clamp Home Sensor Adjustment - Agilent Technologies U1840A Medalist sj5000 Hardware Manual

Automated optical inspection system
Table of Contents

Advertisement

6
Auto Width Adjust Conveyor Setup

Clamp Home Sensor Adjustment

6-16
The Clamp Home sensor has to be adjusted its fit and flatness
between the board and top finger clamp. Follow the steps below
to adjust the sensor.
The tools required are:
• Clamping tolerance test board
• 50 μm filler gauge
• Metric Allen keys
The board clamping plate is 510 mm in length. It is important to
ensure the flatness of the plate along its length when a
production board is clamped.
1
Activate Maintenance mode and manually insert the test
board onto the clamping location.
Click the clamps Up button in the Maintenance window to
2
clamp the board.
3
Insert the 50 μm filler gauge into every clamping tooth that is
in contact with the board. To pass the test, the filler gauge
should bend when you try to insert it between each clamping
tooth and the board.
4
If the clamping mechanism fails the filler gauge test, adjust
the Clamp Home sensor to enable the clamp plate to move
further up and provide a firmer clamp. To do so, loosen the
sensor screws, push up the sensor slightly and tighten the
screws.