4
Starting Up the Medalist sj5000
Before Powering Up
4-2
Before turning the Medalist sj5000 on at the mains isolator, do
the following:
1
Make sure the mains isolator is switched off, then remove the
rear access panel.
2
Measure the voltage between live (phase 1) and neutral
(phase 2) at the mains isolator. The voltage should be 100 to
120 VAC or 200 V to 240 VAC.
3
Ensure the system is correctly earthed.
Now you can power up the sj5000 as described in the following
pages. Refer to
Figure 4-1
buttons to use.
Figure 4-1
Controls on the switch plate at the front of the machine
for the locations of the switches and
Agilent U1840A Medalist sj5000 Hardware Manual