Calibration on Location
Agilent U1840A Medalist sj5000 Hardware Manual
Auto calibration results in new calibrated values which are
saved in CalibratedLightingValues.xml. The values are also
displayed in the Achieved Targets column in the Lighting
Controller window.
If any errors occur during calibration, troubleshoot with the
help of the log file CalibrationResults.log and perform the auto
calibration again.
The following procedure performs automatic calibration at a
specific location.
From the Tools menu, click Lighting Controller under the Camera
1
folder.
Click Calibrate on Location.
2
3
Insert the gray card into the sj5000 and jog the camera over
it. Click Calibrate.
Calibration takes place at the current location. When
calibration is complete, the new calibrated values are saved
and also displayed in the Achieved Targets column in the
Lighting Controller window.
Camera and Lighting Calibration
7
7-13