Lighting Controller Testing
Agilent U1840A Medalist sj5000 Hardware Manual
To test the lighting controller (SLC-800), follow the steps below:
1
In Windows Explorer, navigate to C:\cpi\plc\terminal and
double-click terminal.exe.
1
From the Terminal window, open slc800-new.trm.
Click Settings > Communications.
2
Make sure the settings are as follows, then click OK:
Baud Rate
115.2K
Data Bits
8
Stop Bits
1
Parity
None
Flow Control
None
Connector
COM2
Click MAX CURR 1-4 and check that the feedback is as follows:
3
1MC4000
2MC4000
3MC4000
4MC4000
Click CURRENT 1-4 and check that the feedback is as follows:
4
1SC1100
2SC1200
3SC2500
4SC2700
Click LGT ON + TRIG and check that the feedback is ST050TS0.