10
Setting Up Barcode Readers
Set Up Dual Barcode Reader
10-12
Plug the serial cable (p/n5186-4835) into the XP-40 port on the
breakout box as shown below, and plug other end into the
MS-820 barcode reader.
Plug the trigger cable into the XS-42 port on the breakout box.
There are two setup files for dual lane mode. Two barcode
reader are set up as a daisy chain, and each barcode reader is
identified by a different header.
• 820DL00.esp is the master barcode reader with header 00
• 820DL01.esp is the slave barcode reader with header 01
Program the Barcode Readers
Each barcode reader can only be programmed from port XP-40
on the breakout box.
Install the software as described in
Software"
and do the following:
1
Start the ESP software.
Select MS-820 in the Select a model dialog box and click OK.
2
When prompted to connect to the barcode reader, click Yes.
3
Ensure COM port 4 is selected and click Start.
4
Click App Mode.
5
Click File > Open to open the setup file 820DL00.esp.
6
Click Send/Recv > Send to Reader > Send and Save.
7
Serial cable to barcode reader 1
Serial cable to barcode reader 2
Trigger cable
"Install Microscan ESP