Agilent Technologies U1840A Medalist sj5000 Hardware Manual page 85

Automated optical inspection system
Table of Contents

Advertisement

Agilent U1840A Medalist sj5000 Hardware Manual
Under the Camera folder, click Lighting Controller.
3
4
Select the lighting mode by clicking the first button in the
Lighting area. For post-reflow, use Solder Joints which consists
of the 70 and 60 degree rings.
5
Open the front access door.
6
Drag the E-Stop alarm window to one side so that the camera
display is visible.
Turn the MAINT. SWITCH on the switch plate to the ON
7
position.
Click on clamps Up in the Maintenance window to clamp the
8
board.
9
Once the board is clamped, minimize the Maintenance
window.
10
Set the camera lens aperture to F-stop 5.6 (see
Lens
Aperture").
Camera and Lighting Calibration
"Setting the
7
7-3