Powering Up the sj5000 for the First Time
Agilent U1840A Medalist sj5000 Hardware Manual
When powering up the sj5000 for the first time, carry out the
following steps:
1
Ensure that electrical power has been connected by a
suitably qualified person.
2
Ensure that the mains isolator is in the OFF position.
3
Switch on all other MCBs.
4
Verify that all circuit breakers are on.
5
Measure the input voltage and ensure that it is the rated
voltage of the machine.
6
Ensure that the XY table is free to move over its entire
envelope before continuing. You can manually move it
around its enclosure.
7
Ensure that the conveyor is free from obstruction.
Ensure that the MAINT. SWITCH is in the O position (off).
8
9
Ensure that the front access door is fully closed. Pull out all
the E-Stop buttons before continuing.
10
Ensure the rear access panel is locked in position.
11
Turn on the sj5000 at the mains isolator.
Turn the START key switch to the ON position.
12
Ensure that the green RUN light is lit. If the RUN light is not
13
lit, check the input power.
Ensure the RAIL 1 switch is set to INSP (inspect).
14
15
Turn on the UPS (if installed) and turn on the monitor.
Reset the E-Stop circuit by pressing the ES RESET button.
16
17
Turn on the PC.
Starting Up the Medalist sj5000
4
4-3