8
Camera and Lighting Controller Testing
8-10
Click this:
Check that feedback shows this:
45a_d S7
7840@3*
45a_d S8
7880@7*
The Terminal window should now show the following:
Click Level 4 which returns you to the level 1 window.
17
Click off and check that the feedback reads 0000@0*.
18
At this point none of the LEDs should be on. The lighting
controller is now fully tested.
19
Close the Terminal window. When prompted to save changes