Communication Modules Switch and Jumper Settings
C
HASSIS
For a single analog chassis:
497 .01 Analog Chassis 1
497 .05 Hydraulic Control Panel
For multiple analog chassis:
497 .01 Analog Chassis 1
497 .01 Analog Chassis 2
(optional)
497 .01 Analog Chassis 3
(optional)
497 .01 Analog Chassis 4
(optional)
497 .05 Hydraulic Control Panel
* Jumper X5 is used for the RS-232 communications link (between the first chassis and the test
processor) and the RS-422 communications link (between the 497 communications modules in
each chassis).
† The address switch on the hydraulic control panel should be set to the last number in the chassis
chain. For example, if you have a single 497.01 Analog Chassis, you should set the 497.05 Hydraulic
Control Panel address to 2; If you have two analog chassis, you should set it to 3, and so forth.
Jumper and switch
locations
Communication
Module Jumper
and Switch
Locations
MTS 793 Controller Hardware
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
X1 S
(S3)
WITCH
Set to 1
Set to 2
Set to 1
Set to 2
Set to 3
Set to 4
†
Set to 5
The following graphic illustrates the jumper locations on the 497.36
Communication Module.
3
X6
1
X4
14
8
X10
S4
1
7
S3
S3
X1
Installing 497/498 Electronics (FTIIm)
J
X1, X2, X3
UMPERS
Remove
Install
Remove
Remove
Remove
Remove
Install
X1
X5
4
1
3
2
*
J
X5
UMPER
Install 3 and 4
Install 1 and 2
Install 3 and 4
Install 3 and 4
Install 3 and 4
Install 3 and 4
Install 1 and 2
X3
X2
Installation
67
Need help?
Do you have a question about the FlexTest IIm and is the answer not in the manual?