Cable specification
106
FlexTest GT Controller Connections
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The cable specifications apply to both the analog inputs and analog
outputs.
The cable from the front panel of the I/O Carrier module has the
following specification:
15-contact type D male EMI connector.
Backshell–EMI metallized plastic.
Cable type–up to 6 shielded twisted pairs, each with the drain
wire connected to the signal source.
The cable to the Ch 1 - Ch 6 connectors on the rear panel have the
following specifications:
BNC connector UG88/U
Cable–RG-58 coaxial
Use of a smaller coaxial cable and a RG-174 BNC connector with
an appropriate cable end is permissible.