Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
However, once the buffer has been fully filled, you can retrieve all the stored readings again by
sending this command. For example, if DATA:DATA? "READ, SOURce" is sent, the subsequent
DATA:DATA? "MODE" command will not return mode information. You can get additional buffer data
by sending the command DATA:DATA:SELected?.
When specifying buffer elements, you can:
Specify buffer elements in any order.
Include up to 10 elements in a single list.
Repeat elements as long as the number of elements in the list is less than 10.
Use a comma to delineate multiple elements for a data point.
Use single or double quotes to enclose all the elements.
The options for <bufferElements> are described in the following table.
Option
READing
SOURce
UNIT
MODE
DATE
TIME
TSTamp
RELative
RSTamp
RNUMber
The RELative and RSTamp elements are defined as the time since the first reading was stored in
the buffer. These values are reset every 24-hours. The DATE, TIME, and TSTamp elements
represent absolute time from the internal real-time clock (RTC).
077114601 / March 2019
Description
The measurement reading based on the
measurement function.
The source value.
The unit of measure reading, source value, and
timestamp.
The output state.
The date for the data point.
The time for the data point.
The timestamp for the data point.
The relative time when the data point was measured,
formatted in seconds.
The relative time when the data point was measured.
The reading number indicates the point index in the
trace buffer.
Section 7: SCPI command reference
Example
+8.034562E-03
+4.566723E+01
V, A, and s
CC, CV or OFF
07/01/2013
19:21:36.2556
07/01/2013
19:21:36.2556
+4.430876E+00s
12:23:32.2349
0001
7-179