Section 7: SCPI command reference
:SENSe[n]:<function>:AVERage:WINDow
This command specifies a filter window to control filter threshold.
Type
Command and query
Usage
:SENSe[n]:<function>:AVERage:WINDow <value>
:SENSe[n]:<function>:AVERage:WINDow?
:SENSe[n]:<function>:AVERage:WINDow? DEFault
:SENSe[n]:<function>:AVERage:WINDow? MAXimum
:SENSe[n]:<function>:AVERage:WINDow? MINimum
[n]
<function>
<value>
Details
The <value> parameter can be set to any value from 0.01 to 100. The instrument rounds up the
value to the nearest window value. For example, the filter window is set to 1 if you send the following
command:
SENS:CONC:AVER:WIND 0.5
Example
SENS:CONC:AVER:WIND 0.1
Also see
:SENSe[n]:<function>:AVERage:COUNt
:SENSe[n]:<function>:AVERage[:STATe]
:SENSe[n]:<function>:AVERage:TCONtrol
7-118
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
The measurement function to which this setting applies:
Concurrent: CONCurrent[:DC]
Current: CURRent[:DC]
Voltage: VOLTage[:DC]
The filter window: 0.01 to 100. This value is expressed in terms of a percentage.
There are five acceptable window percentages: 0.01, 0.1, 1, 10, and 100.
(on page 7-115)
(on page 7-116)
(on page 7-117)
Where saved
Save settings
Set the average filter window as 0.1%.
Default value
1
077114601 / March 2019