Section 7: SCPI command reference

:CALCulate2:DATA?

This command reads the results of specified statistics.
Type
Query only
Usage
:CALCulate2:DATA?
:CALCulate2:DATA? <channel>
<channel>
Details
Return the calculation results of statistics. You can get the results in the following order:
Minimum (MINimum)
Maximum (MAXimum)
Mean (MEAN)
Standard deviation (SDEViation)
Peak to peak (PKPK)
If the specified statistics have not been performed yet, the instrument will return -1.000000E+37.
If the buffer is empty, the instrument returns -1.000000E+37.
If there are a lot of readings stored in the buffer, some statistic operations may take too long and
cause a buffer time-out error. To avoid this, send the CALC2:DATA? command and then wait for the
message available (MAV) bit in the Status Byte Register to set before addressing the 2281S to talk.
7-82
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Affected by
Not applicable
The channel number (default value is CH1):
Channel 1: CH1
Channel 2: CH2
Where saved
Not applicable
Default value
Not applicable