Output:protection:clear - Tektronix Keithley 2281 Series Reference Manual

Precision dc power supply and battery simulator
Table of Contents

Advertisement

Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual

:OUTPut:PROTection:CLEar

This command clears the overvoltage, overcurrent, overtemperature, or reverse fault status.
Type
Command only
Usage
:OUTPut:PROTection:CLEar
:OUTPut:PROTection:CLEar <channel>
<channel>
Details
Clears the protection status if an overvoltage, overcurrent, or overtemperature protection condition
occurred or a reverse sense fault was detected. For all four faults, this command also clears the
related Questionable Instrument Summary Condition Register bit.
Example
:OUTP:PROT:CLE
Also see
:OUTPut:PROTection:TRIPped?
Status model
077114601 / March 2019
Affected by
Not applicable
The channel number (default value is CH1):
Channel 2: CH2
(on page 7-112)
(on page D-1)
Where saved
Not applicable
Channel 1: CH1
Clear the protection status.
Section 7: SCPI command reference
Default value
Not applicable