Section 7: SCPI command reference
:STATus:QUEStionable:ENABle
This command sets or reads the contents of the Questionable Event Enable Register of the status model.
Type
Command and query
Usage
:STATus:QUEStionable:ENABle <n>
:STATus:QUEStionable:ENABle?
[n]
Details
This command sets or reads the contents of the enable register of the Questionable Event Register.
When one of these bits is set and the corresponding bit in the Questionable Event Register is set, the
QSM bit in the Status Byte Register is set.
When sending binary values, preface <n> with #b. When sending hexadecimal values, preface <n>
with #h. No preface is needed when sending decimal values.
Example
:STAT:QUES:ENAB #b0010000000000000
Also see
Questionable Event Register
:STATus:QUEStionable:INSTrument[:EVENt]?
This command reads and clears the Questionable Instrument Event Register of the status model.
Type
Query only
Usage
:STATus:QUEStionable:INSTrument[:EVENt]?
Details
This attribute reads the Questionable Instrument Event Register of the status model and then clears
the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see
Also see
None
7-158
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Affected by
Power cycle
The value of the Questionable Status Enable Register
(on page D-14)
Affected by
Not applicable
Questionable Instrument Event Register
Where saved
Not applicable
Sets the operation Status Enable Register
using a binary value.
You could also send the decimal value 8192
or the hexadecimal value #h2000.
Where saved
Not applicable
(on page D-15).
Default value
0
Default value
Not applicable
077114601 / March 2019
Need help?
Do you have a question about the Keithley 2281 Series and is the answer not in the manual?