Arm[:Sequence[N]]:Source - Tektronix Keithley 2281 Series Reference Manual

Precision dc power supply and battery simulator
Table of Contents

Advertisement

Section 7: SCPI command reference

:ARM[:SEQuence[n]]:SOURce

This command selects the control source for the arm layer.
Type
Command and query
Usage
:ARM[:SEQuence[n]]:SOURce <source>
:ARM[:SEQuence[n]]:SOURce?
[n]
<source>
Example
ARM:SOUR BUS
Also see
:TRIGger[:SEQuence[n]]:SOURce
Trigger model
7-22
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
The trigger source for the arm layer:
Immediate triggering; no delay or gating in arm detection layer:
IMMediate.
External triggering: EXTernal
Manual triggering; the trigger signal is generated by pressing the
TRIGGER key: MANual
Bus triggering by the command *TRG: BUS
(on page 7-200)
(on page 3-30)
Where saved
Save settings
Select bus triggering as the trigger source
for the arm layer.
Default value
IMM
077114601 / March 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Keithley 2281 Series and is the answer not in the manual?

This manual is also suitable for:

Keithley 2281s-20-6

Table of Contents