Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Battery test measure reset values
Setting*
MENU > Measure > V & I > Sample Interval
:BATTery[n]:TEST:SENSe:SAMPle:INTerval
(on page 7-57)
MENU > Measure > Voc / ESR > Delay
:BATTery[n]:TEST:SENSe:EVOC:DELAy
page 7-56)
MENU > Measure > A-H > V-Full
:BATTery[n]:TEST:SENSe:AH:VFULl
7-56)
MENU > Measure > A-H > Source I-Limit
:BATTery[n]:TEST:SENSe:AH:ILIMit
7-55)
MENU > Measure > A-H > ESR-Interval
:BATTery[n]:TEST:SENSe:AH:ESRInterval
page 7-51)
MENU > Measure > V & I > Sample Interval
:BATTery[n]:TEST:SENSe:SAMPle:INTerval
(on page 7-57)
* Some of the options listed in this table are only available for specific functions (power supply, battery test, or battery
simulation). If you do not see the option listed here on the menu, it may not be available for the function you are using.
077114601 / March 2019
Default value reset from
MENU > System > Settings
> Reset
0.04 s for 50 Hz
33.333 ms for 60 Hz
100.00 ms
(on
0.000 V
(on page
0.1 A
(on page
10 s
(on
0.04 s for 50 Hz
33.333 ms for 60 Hz
0.04s for
50Hz3.33333e-2s for 60Hz
0.04s for 50Hz
Section 2: General operation
Default value reset by the
*RST
(on page 7-7)
command or
:SYSTem:PRESet
(on
page 7-175) command
0.04 s for 50 Hz
3.33333e-02 s for 60 Hz
+1.00000e-01
0.0
0.1000
10 s
0.04 s for 50 Hz
3.33333e-02 s for 60 Hz
2-119
Need help?
Do you have a question about the Keithley 2281 Series and is the answer not in the manual?