Initiate Subsystem; Initiate[N]:Continuous - Tektronix Keithley 2281 Series Reference Manual

Precision dc power supply and battery simulator
Table of Contents

Advertisement

Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual

INITiate subsystem

The commands in the INITiate subsystem control the trigger model.

:INITiate[n]:CONTinuous

This command enables and disables continuous triggering.
Type
Command and query
Usage
:INITiate[n]:CONTinuous <state>
:INITiate[n]:CONTinuous?
[n]
<state>
Details
With continuous initiation enabled, sending :INITiate causes error -213, "init ignored."
Example
ARM:SOUR IMM
TRIG:SOUR IMM
INIT:CONT OFF
ABORt
ARM:COUN 1
TRIG:COUN 1
TRIG:SAMP:COUN 100
:OUTP ON
INIT
Also see
Trigger model
077114601 / March 2019
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
Disable continuous triggering: OFF or 0
Enable continuous triggering: ON or 1
(on page 3-30)
Section 7: SCPI command reference
Where saved
Save settings
Select IMMediate arm source.
Select IMMediate trigger source.
Disable continuous initiation.
Abort operation. Places the instrument in idle.
Set arm count to 1.
Set trigger count to 1.
Set Measure count to 100.
Turn on the output.
Start measurement process.
Default value
0 (OFF)
7-105

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Keithley 2281 Series and is the answer not in the manual?

This manual is also suitable for:

Keithley 2281s-20-6

Table of Contents