Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
:BATTery[n]:TEST:MEASure:EVOC?
This command queries the values of ESR and Voc when the power supply is in Battery Test mode.
Type
Query only
Usage
:BATTery[n]:SIMulator:MEASure:EVOC?
[n]
Details
When this command is executed, the instrument starts a new measurement and returns the values of
ESR and Voc. Output is off after the measurement.
Example
:BATT:TEST:MEAS:EVOC?
Also see
:BATTery[n]:TEST:SENSe:EVOC:DELAy
:BATTery[n]:TEST:SENSe:AH:ESRInterval
This command sets and queries the ESR sample interval during A-H measurement in the battery test function.
Type
Command and query
Usage
:BATTery[n]:TEST:SENSe:AH:ESRInterval <interval>
:BATTery[n]:TEST:SENSe:AH:ESRInterval?
[n]
<interval>
Example
:BATT:TEST:SENS:AH:ESRI S60
:BATT:TEST:SENS:AH:ESRI?
Also see
None
077114601 / March 2019
Affected by
Not applicable
Channel number: For single-channel instruments, the value is 1
(on page 7-56)
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
S10 = 10 seconds
S30 = 30 seconds
S60 = 60 seconds
S120 = 120 seconds
M10 = 10 minutes
Section 7: SCPI command reference
Where saved
Not applicable
The values of ESR and Voc are returned.
Where saved
Save settings
Set the ESR sample interval to 60 s and query the
value.
Return: S60
Default value
Not applicable
Default value
S10
7-51
Need help?