Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
[:SOURce[n]]:LIST:END:ZERO
This command specifies the source values when the instrument is stopping a list.
Type
Command and query
Usage
[:SOURce[n]]:LIST:END:ZERO <value>
[:SOURce[n]]:LIST:END:ZERO?
[n]
<value>
Details
This command allows you to end a specified list with the last source value or zero value.
Example
LIST:END:ZERO ON
Also see
None
[:SOURce[n]]:LIST:HTIMe
This command determines the hold time of the list running.
Type
Command and query
Usage
[:SOURce[n]]:LIST:HTIMe <type>
[:SOURce[n]]:LIST:HTIMe?
[n]
<type>
Hold time is decided by whichever is longer: dwell time or sample count.
077114601 / March 2019
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1.
When the list is stopped,
Use the last source values: OFF or 0
Set the voltage to 0 V, current source to 0.1 A: ON or 1
Affected by
Power cycle
Channel number: For single-channel instruments, the value is 1
POINt: The list runs until the reading reaches the preset sample count; the hold time
is computed by sample count * sample speed; the preset dwell time is ignored in
this mode
TIME: The list runs until the preset time is reached; the preset sample count is
ignored in this mode
Section 7: SCPI command reference
Where saved
Save settings
List ends with the 0 V and 0.1000 A.
Where saved
Not applicable
Default value
0 (OFF)
Default value
POINt