Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Example
:BATT:TEST:SENS:EVOC:DEL 0.05
Also see
:BATTery[n]:TEST:MEASure:EVOC?
:BATTery[n]:TEST:SENSe:SAMPle:INTerval
This command sets and queries the sample interval in the battery test function.
Type
Command and query
Usage
:BATTery[n]:TEST:SENSe:SAMPle:INTerval <value>
:BATTery[n]:TEST:SENSe:SAMPle:INTerval?
:BATTery[n]:TEST:SENSe:SAMPle:INTerval? DEFault
:BATTery[n]:TEST:SENSe:SAMPle:INTerval? MINimum
:BATTery[n]:TEST:SENSe:SAMPle:INTerval? MAXimum
[n]
<value>
Details
The sample interval is based on the power line cycles.
The shorter sample interval with the faster reading rate would increase reading noise and decrease
the number of usable digits, while the longer sample interval with the lower reading rate would provide
lower reading noise and more readable digits.
Example
:BATT:TEST:SENS:SAMP:INT 0.1
Also see
None
077114601 / March 2019
(on page 7-51)
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
8e-5 to 0.48s for 50 Hz
6.66667e-5 to 0.5s for 60 Hz
Section 7: SCPI command reference
Set the EVoc delay to 0.05 s.
Where saved
Save settings
Set the sample interval to 0.1 s.
Default value
0.04s for 50 Hz
3.33333e-2s for 60 Hz
7-57