Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Details
There are two types of instrument commands:
Overlapped commands: Commands that allow the execution of subsequent commands while
instrument operations of the overlapped command are still in progress.
Sequential commands: Commands whose operations must finish before the next command is
executed.
The *WAI command suspends the execution of commands until the instrument operations of all
previous overlapped commands are finished. The *WAI command is not needed for sequential
commands.
The 2281S has the following overlapped commands:
*TRG
:INITiate
:CALibration:PROTection:STEP1:DATA
:CALibration:PROTection:STEP2:DATA
:CALibration:PROTection:STEP3:DATA
:CALibration:PROTection:STEP4:DATA
:CALibration:PROTection:STEP5:DATA
:CALibration:PROTection:STEP6:DATA
:CALibration:PROTection:STEP7:DATA
:CALibration:PROTection:STEP8:DATA
Also see
CALibration subsystem
*OPC
(on page 7-6)
*TRG
(on page 7-9)
077114601 / March 2019
(on page 7-90)
Section 7: SCPI command reference
7-11