Battery[N]:Simulator:voc:full - Tektronix Keithley 2281 Series Reference Manual

Precision dc power supply and battery simulator
Table of Contents

Advertisement

Section 7: SCPI command reference

:BATTery[n]:SIMulator:VOC:FULL

This command sets the threshold Voc value for the charging to stop in the battery simulator function.
Type
Command and query
Usage
:BATTery[n]:SIMulator:VOC:FULL <value>
:BATTery[n]:SIMulator:VOC:FULL?
:BATTery[n]:SIMulator:VOC:FULL? DEFault
:BATTery[n]:SIMulator:VOC:FULL? MINimum
:BATTery[n]:SIMulator:VOC:FULL? MAXimum
[n]
<value>
Details
This command sets the threshold Voc value for the charging to stop in the battery simulator function.
If the full Voc is smaller than the empty Voc, the battery model cannot be accessed and you will see
error 703, "Full Voc can't be less than empty Voc."
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.
Example
:BATT:SIM:VOC:FULL 5
Also see
:BATTery[n]:SIMulator:VOC:EMPTy
7-50
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
The value range of the Voc: Minimum to Maximum
(on page 7-49)
Where saved
Save settings
Set the full Voc value to 5 V.
Default value
Maximum
077114601 / March 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Keithley 2281 Series and is the answer not in the manual?

This manual is also suitable for:

Keithley 2281s-20-6

Table of Contents