Section 2: General operation
To clear the error, you need to clear the Questionable Instrument Summary Event Register bit and
initiate the trigger model using the front panel or SCPI commands. For more information on registers,
refer to
Questionable Instrument Summary Event Register
To clear an overprotection error using the front panel:
When the overprotection error dialog box is displayed, you can press the ENTER key to initiate the
trigger model.
To clear an overprotection error using a SCPI command:
When an overprotection error occurs, send the following command to clear the register and initiate
the trigger model:
:OUTPut:PROTection:CLEar

Remote communications interfaces

You can choose from one of several communication interfaces to send commands to and receive
responses from the 2281S.
The 2281S can be controlled from only one communication interface at a time. The first interface from
which it receives a message takes control of the instrument. To remove control from the present
interface, send the following command:
:SYSTem:COMMunication:ABORt
The 2281S powers up using the last remote interface used.
2-68
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Figure 42: Reversed sense leads error
(on page D-16).
077114601 / March 2019