[:Source[N]]:List:save:internal - Tektronix Keithley 2281 Series Reference Manual

Precision dc power supply and battery simulator
Table of Contents

Advertisement

Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual

[:SOURce[n]]:LIST:SAVE:INTernal

This command stores the specified list to nonvolatile memory.
Type
Command only
Usage
[:SOURce[n]]:LIST:SAVE:INTernal <listNumber>
[n]
<listNumber>
Details
The 2281S supports up to nine lists. You can save a specified list to internal memory. Before saving,
the instrument will verify the points of voltage, current, and dwell in the specified list. If they are the
same, the list will be saved. Otherwise, error 202, "List lengths not equivalent," is generated.
If a list already exists for the specified list number, the new list overwrites the existing list.
Example
LIST1:VOLT "1,2"
LIST1:CURR "0.5,1"
LIST1:DWEL "0.5,0.5"
LIST:SAVE:INT 1
Also see
[:SOURce[n]]:LIST<listNumber>:<element>
[:SOURce[n]]:LIST<listNumber>:<element>:APPend
077114601 / March 2019
Affected by
Not applicable
Channel number: For single-channel instruments, the value is 1
The list number: 1 to 9
(on page 7-134)
Section 7: SCPI command reference
Where saved
Not applicable
Create a list with 2 points:
Point 1: 1 V, 0.5 A, 0.5 s
Point 2: 2 V, 1 A, 0.5 s
Save list 1 to the internal memory
(on page 7-135)
Default value
Not applicable
7-143