Filling and mounting
D
1 After any cleaning, push the absorber insert
back into the absorber container (D)
completely.
2 Fill CO
absorber with fresh soda lime to the
2
upper mark.
Recommendation: Only use Drägersorb 800 Plus
or Drägersorb Free.
WARNING
Risk of hypoventilation and incorrect gas
measurement
Reuse of the disposable dust filter can
increase the filter resistance and impair the
ventilation function of the device.
If the reusable CO
absorber is used, always
2
use a disposable dust filter. Replace the
disposable dust filter with every change of
soda lime.
E
Instructions for use Perseus A500 SW 2.0n
F
3 Insert a new disposable dust filter (E). Only use
dust filters from the list of accessories. Only
use undamaged filters.
4 Insert CO
absorber (F) on the breathing
2
system from below and then rotate it
counterclockwise as far as it will go.
Strictly observe the instructions for use for
Drägersorb 800 Plus or Drägersorb Free soda
lime.

Water trap

 Empty or replace the water trap according to its
instructions for use.
WARNING
Risk due to full water trap
When the water trap is full, the gas
measurement can fail and insufficient
ventilation may occur.
Check the water level in the water trap
regularly and empty it as necessary.
CAUTION
Risk due to gas measurement failure
Contaminants, damage, or overfilling of the water
trap can influence gas measurement.
– Observe the instructions for use of the water
trap.
– Always operate the device with a mounted
water trap.
WARNING
Risk of infection
The water trap may contain infectious fluid.
– Proceed carefully when emptying and take
protective measures if necessary.
– Follow the infection prevention policies
and reprocessing regulations of the
healthcare facility.
Assembly and preparation
77