Adopting alarm settings when changing the ventilation mode

When the ventilation mode is changed, the alarm
settings are also adapted.
Depending on the mode, alarm settings can either
be adopted or set to Off.
Alarm or alarm
limit
PC - APRV (optional)
PSV (optional) with
FiO
low
2
inAgent high
Apnea (no CO
)
2
etCO
high
2
etCO
low
2
inCO
high
2
FiO
high
2
Is restored or remains
inAgent low
xMAC low
Paw high
Paw low
MV high
MV low
1)
Apnea (no flow)
Apnea (no pres-
sure)
1) This alarm is only activated when the MV low alarm limit is also activated.
Instructions for use Perseus A500 SW 2.0n
VC,
Pause, Man/Spon,
VC - AF,
PSV (optional) with
PC,
∆Psupp <5 hPa
∆Psupp
5 hPa
active
Alarm settings are
configurable, see
On
Some modes can be configured whether or not the
settings are adopted.
However, the settings can be adjusted at any time
during operation.
Mode
Ext. FGO
(optional)
Alarm settings are adopted
Alarm settings are adopted
configurable, see page 156
adopted
Not mea-
page 156
CBM mode
Off
configurable, see
page 156
Not activated
Alarm settings are
adopted
sured
configurable, see
page 156
Off
Alarms
143