Inspection; Remote Service; Safety Checks - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Inspection

Measure
Inspection and safety check

Remote service

Perseus supports the following Remote Service
functions:
– Help ticket
– Remote Device Check
– Proactive call
Contact the responsible DrägerService
organization for further information on the Remote
Service function.

Safety checks

Safety checks are no substitute for preventive
maintenance measures (including preventive
replacement of wear parts) as identified by the
manufacturer.
Performing the safety checks
1 Check whether the corresponding instructions
for use are available in the correct version.
2 Perform a functional test of the following
functions in accordance with the instructions
for use:
– Emergency O
2
3 Check that the product is in good condition:
– All labels are complete and legible
– No visible damage to the following
components:
– Trolley and castors
– Housing parts
– Brakes
– Auxiliary power sockets
– Vaporizer mount
– Water trap mount
– O-rings of the water trap mount
Instructions for use Perseus A500 SW 2.0n
delivery
Interval
Every 12 months
– Screen
– Gas inlets
– Status display
– Breathing system
– AGS and AGS valves
– Hoses and cables
– Strain reliefs for AGS, compressed gas
hoses, cables
– Optional: Holding arms
– Optional: Flexibility trolley, castors, and
locking mechanisms
– Optional: Pressure reducers and their
sensor lines
– Fuses which are accessible from the
outside are in compliance with the specified
values
– Check that the country-specific labeling of
the gas type matches the screen display.
– Check the end stop of the holding arms for
patient monitors.
4 Check the electrical safety in accordance with
the IEC 62353 standard.
5 Check the safety equipment:
– Functional integrity of optical and acoustic
alarm generators
– O
switch on gas mixing unit (electronically
2
controlled gas mixer)
– Internal battery
– Optional: Check the function of the
minimum O
controlled gas mixer).
User group
Service personnel
delivery (mechanically
2
Service
217