Dräger Perseus A500 Instructions For Use Manual page 146

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Alarms
Alarm
Airway press. continuously high
Airway pressure negative
PEEP/CPAP high
Airway pressure not achieved
Cardiac bypass mode still
active?
Fresh gas low or leakage
O
measurement not available
2
N
O measurement not available
2
Agent measurement not avail-
able
O
measurement temporarily
2
inaccurate
CO
sensor accuracy low
2
Agent measurement temporar-
ily inaccurate
N
O measurement temporarily
2
inaccurate
Measured gas concentrations
temporarily inaccurate
146
Low
---
---
---
---
---
Breathing bag almost
empty
>20 s
Priority
Medium
---
---
Airway pressure
>(PEEP +5 hPa)
during more than 10
consecutive breaths
>15 seconds (for RR
6)
or
>30 s
(for RR <6 or PSV)
If a minute volume of
>50 % of the sug-
gested value is mea-
sured after CBM
mode has been acti-
vated for >60 seconds
Breathing bag empty
---
Instructions for use Perseus A500 SW 2.0n
High
> 15 seconds above
the manually or auto-
matically set limit
Pmean < –2
or
Paw < –10
---
---
---
after 30 s
or
in the event of the
additional alarm
"Apnea (no flow)" or
"Apnea (no pressure)"
or
in the event of the
additional alarm
"Emergency air inlet
activated"
---