Operation
Only gas concentration
measurement/use of Monitoring
No ventilation in Monitoring operation mode. Gas
delivery is also stopped with electronically
controlled gas mixers. The gas concentration
measurement remains active and waiting for
respiratory phases.
Use this operating mode for regional anesthesia,
for example.
Prerequisite: The device is in Standby mode or
Man / Spon mode.
Activating
 Start Monitoring operating mode.
Deactivating
 After Standby or change to ventilation mode.
For more information, see page 162.
Pausing the therapy/using Pause
(optional)
In Pause mode, ventilation is stopped. Gas
delivery is also stopped with electronically
controlled gas mixers. The gas concentration
measurement remains active and waiting for
respiratory phases.
This mode is equipped with an adjustable therapy
control Timer, depending on the patient category.
When the set time has elapsed, this emits an
alarm as reminder that ventilation should be
resumed. There is no automatic resumption of
ventilation. Setting the therapy control Timer to
"Off" deactivates the alarm. The total elapsed time
is also displayed.
Use this operating mode for regional anesthesia or
short pauses in therapy, such as disconnection or
intubation, to reduce contamination of the ambient
air with anesthetic gases due to an open Y-piece.
94
Activating
1 Start Pause operating mode.
2 Adjust the therapy control Timer if necessary.
Returning to the previous mode
1 Touch the Resume ventilation button.
2 Confirm the ventilation mode.
For more information, see page 155 and page 162.
Instructions for use Perseus A500 SW 2.0n