Configuration
Vertical tab "System test"
Headline/
Parameter
General
Always use walk-
through mode
Sample line is con-
nected during test
Test gas supply
Central O
supply
2
Central Air supply
Central N
O supply
2
O
cylinder
2
Air cylinder
N
O cylinder
2
Verify O
delivery
2
166
Setting range
On; Off
On; Off
On; Off
On; Off
On; Off
On; Off
On; Off
On; Off
On; Off
Description
When this function is switched on,
system tests and leakage tests will
always be executed in walk-
through mode.
Specify whether the sample line is
connected to the Y-piece or to the
filter on the Y-piece during the
automatic tests.
When On is configured, the device
can automatically check whether,
e.g., breathing hoses or central
supply hoses are incorrectly con-
nected.
When Off is configured, additional
checks are required during the
system test.
Select which gas supplies are
tested during the automatic sys-
tem test.
If Auto On is configured, only
check the gas supplies whose
actual availability is guaranteed
after the automatic start-up. Set-
ting both O
sources to Off - in
2
order to make the system test suc-
ceed - is not possible.
Select whether the system test will
check that the O
ally delivering oxygen.
The Sample line is connected
during test parameter must also be
set to On for this.
Instructions for use Perseus A500 SW 2.0n
supply is actu-
2