Leakage Assistant - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Tests
After the test, the final test result is displayed on
the Standby page, see page 81.
Test interruption due to irregularities
If an irregularity is detected during the automatic
test, the following occurs:
– The test is interrupted.
– An acoustic signal sounds.
– Information on the cause and remedy are
displayed.
Test details
Remedying the cause:
1 Remedy the cause of the interruption.
If there is leakage, the leakage assistant (K)
can be used to support troubleshooting.
2 Touch the Repeat button (I) and repeat the test
of the component.
Accepting the irregularity:
 Touch the Accept button (J) and continue the
test.
Accepted irregularities prevent the total result of
the system test from indicating "fully operational"
and are protocoled in the logbook.
124

Leakage assistant

Prerequisite when vaporizers are connected:
– Vaporizer is vertically level and securely
mounted on the plug-in adapter.
– Filling inlet is closed.
In this test, a continuous pressure is generated
and the current leakage value is displayed.
Leakage assistant
K
I
If the leakage value changes as a result of
J
changes made to the device (e.g., loosening or
readjusting hose connections), this can help locate
the cause of the leakage.
The displayed leakage value may differ from the
value that was determined in the leakage test. The
reasons for this are the different measuring
methods and different pneumatic ranges.
1 Touch the button for the test (Leakage
assistant).
2 Follow the instructions (A).
Checking a vaporizer for leakage
Prerequisites:
– Gas mixer:
– With electronically controlled gas mixer:
The flow control valve of the O
is closed.
– With mechanically controlled gas mixer:
All flow control valves are closed.
– The vaporizers are closed.
– The breathing circuit is correctly connected.
– The APL valve is set to 30.
Instructions for use Perseus A500 SW 2.0n
A
flowmeter
2