Advertisement

Vertical tab "Units"
Headline/
Parameter
Weight
Height
Airway pressure
Supply pressure
CO
2
Volatile agents
Vertical tab "Auto On/Flush & Dry"
Headline/
Parameter
Auto On
Day and time
Monday to Sunday; Hour : Minute
Monday to Friday: 6:30
Saturday & Sunday: Off
When completed
Standby/ Shut down
Flush & Dry
Flush duration
between cases
[min]
Flush & Dry dura-
tion before shut-
down [min]
Instructions for use Perseus A500 SW 2.0n
Setting range
kg; lbs
cm; in
mbar; hPa; cmH
O
2
bar; kPa×100; psi
%; kPa; mmHg
%; kPa
Setting range
Off; 1 to 5
2
Off; 5 to 30
10
Configuration
Description
Set units.
Description
Set the Auto On option. Perseus
provides a calendar function
which, when the device is shut
down, suggests the next time the
device should be ready for use,
depending on the weekday. To
activate the automatic ready-for-
use time for any desired day,
select the corresponding day and
set the time, see page 136.
Pay attention to the correct setting
of the gas supplies to be tested,
see page 166.
Set the flush cycle duration, e.g.,
at patient change.
Off:
Flush function is deactivated.
Set the duration of flushing and
drying when the device is shut
down.
Off:
Drying function is deactivated.