Selecting And Connecting Patient-Specific Accessories; Mounting The Flexible Arm (Optional) Or The Bag Elbow; Safety Information - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Mounting the flexible arm (optional) or
the bag elbow
The breathing bag can be mounted either on the
flexible arm or, using the bag elbow and a
breathing hose, mounted directly on the breathing
system.
Mounting the flexible arm
C
D

Selecting and connecting patient-specific accessories

Safety information

WARNING
Risk due to leakage in coaxial breathing
circuit
Leakage in the inner hose of a coaxial
breathing circuit may result in CO
rebreathing or inadequate gas exchange. The
device can only detect such leakage if a
separate test with a coaxial test adapter is
performed.
– Use a coaxial test adapter to check the
inner hose for leakage, see page 125.
Next, perform a leakage test on the entire
breathing circuit.
– Monitor the measured gas concentrations
during ventilation.
Instructions for use Perseus A500 SW 2.0n
A
B
2
1 Attach the attachment piece of the arm (B) to
the connection piece (A) on the breathing
system and screw down tightly with the two
knurled screws. Check that the arm is fixed
securely!
2 Attach the elbow (C) to the end of the flexible
arm.
3 Align the flexible arm so that collisions with
other mounted assemblies are prevented.
Mounting the bag elbow
 Attach the bag elbow (D) directly to the
connection piece on the breathing system.
NOTE
This device is made without natural rubber latex.
To minimize the risk of exposure to latex, use
breathing bags and breathing hoses that are not
made with natural rubber latex.
Assembly and preparation
67

Advertisement

Table of Contents
loading

Table of Contents