Device outputs (continued)
Pin assignment
Pin 1
Pin 2
Pin 3
Pins 4, 6
Pin 5
Pins 7, 8
Pin 9
Housing
USB interface
Type
Supported devices
LAN interface
Type
Transfer speed
Connection for external workstation light
(optional)
Distribution piece for central O
(optional)
Supply pressure
Maximum permissible extraction flow rate
Instructions for use Perseus A500 SW 2.0n
n/c
RXD
TXD
Pins 4 and 6 are connected internally
SHLD-GND
Pin 7 and Pin 8 are connected internally
n/c
SHLD-GND
Only connect USB storage media that do not
have their own power supply. Do not connect
any charging cables.
USB type A connector; USB 1.1
USB flash drive formatted with FAT16 or FAT32
Dräger recommends storage media with FIPS
140-2 compatible encryption and their own
encryption facility.
Only for Dräger Remote Service
Only connect devices and/or networks that meet
the requirements of IEC 60950-1/IEC 62368-1
for ungrounded SELV circuits and the require-
ments of IEC 60601-1 (as of the 2nd edition) for
touchable secondary circuits with a maximum
nominal voltage of 24 V DC.
RJ45 plug
10Base-T, IEEE 802.3 Clause 14.
Requires at least a CAT3 cable.
Only for workstation lights approved by Dräger,
see list of accessories.
supply
2
2.7 to 6.9 kPa x 100
39 to 100 psi
0.27 to 0.69 MPa
40 L/min
Technical data
245