Priority
Alarm
!
100
Infinity ID CO
not compatible
!
100
Infinity ID water trap not
compatible
!! 100
Insp. press. sensor failure
!!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
!
75
Inspiratory desflurane low
!!! 255
Inspiratory enflurane high
!! 255
Inspiratory enflurane high
!
75
Inspiratory enflurane low
!
100
Inspiratory flow sensor
expired
!
190
Inspiratory flow sensor not
calibrated
Instructions for use Perseus A500 SW 2.0n
Cause
absorber
An incompatible accessory is
2
connected.
An incompatible accessory is
connected.
Sensor calibration failed.
concentration has exceeded
the upper alarm limit.
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Accessory has been used too
long.
The sensor is not calibrated.
The breathing system has
been replaced or discon-
nected since last calibration.
Failure of the flow sensor.
Troubleshooting
Remedy
Check accessory.
Use "ALARM RESET" to
acknowledge alarm.
Check accessory.
Use "ALARM RESET" to
acknowledge alarm.
Change to Man/Spon mode
and ventilate manually.
Perform the system test.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakages.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakages.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
Perform the leakage test.
Replace the flow sensor. Per-
form the leakage test.