Volume-controlled ventilation

Compliance correction
The applied VT is corrected by the determined
breathing hose compliance, i.e., an additional
volume is delivered in order to ensure the
VC
Paw
Ti
Flow
VC - CMV
– Volume-controlled
– Pressure-limited
– Time-controlled
Instructions for use Perseus A500 SW 2.0n
application of the volume to the patient. The
applied VT is also corrected by the sampling flow
of the built-in patient-gas measurement module as
soon as CO
%Tplat
Slope
1/RR
Trigger
– Machine-triggered
– Constant inspiratory flow
respiratory phases are detected.
2
∆Psupp
PEEP
Trigger
Insp. term.
Principles of operation