Dräger Perseus A500 Instructions For Use Manual page 169

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Vertical tab "Infinity ID"
Headline/
Parameter
Monitoring of Infinity ID accessories
Breathing circuit
Water trap
Flow sensors
CO
absorber
2
Exchange interval [days]
Breathing circuit
Water trap
Flow sensors
CO
absorber
2
Vertical tab "Service"
Headline/
Parameter
Instructions for use Perseus A500 SW 2.0n
Setting range
On; Off
On; Off
On; Off
On; Off
Off; 2 to 9
2
Off; 28
Off; 1 to 180
90
Off; 1 to 28
7
Setting range
Configuration
Description
Activate or deactivate the
Infinity ID functionality.
On:
– Generates a message when
the maximum period of use is
exceeded
– Generates a message when
Infinity ID breathing hoses are
incorrectly connected
Off:
– Messages are suppressed.
Set the exchange intervals for
Infinity ID accessories.
Description
The following functions are avail-
able after the appropriate creden-
tials have been entered:
– Access to the service dialog
169

Advertisement

Table of Contents
loading

Table of Contents