Aprv (Optional) - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Principles of operation

APRV (optional)

Paw
Phigh
Plow
Slope
Thigh
Tlow
Flow
PC - APRV
– Pressure-controlled
– Time-controlled
– Machine-triggered
– Spontaneous breathing at continuous positive
airway pressure with short pressure releases
In PC - APRV, the patient's spontaneous breathing
occurs at the upper pressure level Phigh. This
pressure level is maintained for the duration of
Thigh.
The number of pressure releases is determined by
the Thigh and Tlow settings. The releases are
time-controlled and are not triggered by the
patient. The duration is determined by Tlow.